| Outputs | All channel outputs are galvanically isolated and not connected to ground. | 
| Number of output channels | R&S®NGE102B | 2 | 
|  | R&S®NGE103B | 3 | 
| Total output power | R&S®NGE102B | max. 66 W | 
|  | R&S®NGE103B | max. 100 W | 
| Maximum output power | per channel | 33.6 W | 
| Output voltage | per channel | 0 V to 32 V | 
| Maximum output current | per channel | 3 A | 
| Maximum voltage in serial operation | R&S®NGE102B | 64 V | 
|  | R&S®NGE103B | 96 V | 
| Maximum current in parallel operation | R&S®NGE102B | 6 A | 
|  | R&S®NGE103B | 9 A | 
| Voltage ripple and noise | 20 Hz to 20 MHz | typ. < 1.5 mV (RMS) < 20 mV (peak-to-peak) (meas.)
 | 
| Current ripple and noise | 20 Hz to 20 MHz | < 2 mA (RMS) (meas.) | 
| Load regulation | load change from 10% to 90% |  | 
| Voltage | ± (% of output + offset) | < 0.1% + 20 mV | 
| Current | ± (% of output + offset) | < 0.1% + 5 mA | 
| Load recovery time | 10% to 90% load change within a band of ± 30 mV of set voltage | < 200 µs (meas.) | 
| Line regulation | ±10% change in mains voltage |  | 
| Voltage | ± (% of output + offset) | < 0.1% + 20 mV | 
| Current | ± (% of output + offset) | < 0.1% + 5 mA | 
| Output voltage overshoot at turn-off of mains power and active channel output |  | < 100 mV | 
| Programming resolution |  |  | 
| Voltage |  | 10 mV | 
| Current |  | 1 mA | 
| Programming accuracy |  |  | 
| Voltage | ± (% of output + offset) | < 0.1% + 30 mV | 
| Current | ± (% of output + offset) | < 0.1% + 5 mA | 
| Output measurements | 
| Measurement functions |  | voltage, current, power | 
| Readback resolution |  |  | 
| Voltage |  | 10 mV | 
| Current |  | 1 mA | 
| Readback accuracy |  |  | 
| Voltage | ± (% of output + offset) | < 0.1% + 20 mV | 
| Current | ± (% of output + offset) | < 0.1% + 5 mA | 
| Temperature coefficient | +5°C to +20°C and +30°C to +40°C | 
| Voltage | ± (% of output + offset) | < 0.02% + 5 mV per K | 
| Current | ± (% of output + offset) | < 0.02% + 3 mA per K | 
| Ratings | 
| Maximum voltage to earth |  | 150 V DC | 
| Maximum counter voltage | voltage with same polarity connected to the outputs | 33 V | 
| Maximum reverse voltage | voltage with opposite polarity connected to the outputs | 0.4 V | 
| Maximum reverse current |  | 3 A | 
| Remote control | 
| Command processing time |  | ≤ 30 ms | 
| Programming settling time | within 1% of final value |  | 
| Positive voltage change | no load | 10 ms + command processing time | 
|  | full load (resistive) | 10 ms + command processing time | 
| Negative voltage change | no load | 500 ms + command processing time | 
|  | full load (resistive) | 10 ms + command processing time | 
| Protection functions | 
| Overvoltage protection |  | configurable for each channel | 
| Overpower protection |  | configurable for each channel | 
| Overcurrent protection (electronic fuse) |  | configurable for each channel | 
| Response time | (Iload > Iresp × 2) | < 10 ms | 
| Fuse linking (FuseLink function) |  | Yes | 
| Response time of linked channels |  | < 100 µs + response time of linked channel | 
| Fuse delay | configurable for each channel | 10 ms to 10 s (10 ms increments) | 
| Overtemperature protection |  | Yes | 
| Special functions | 
| Output ramp function |  | EasyRamp | 
| EasyRamp time |  | 10 ms to 10 s (10 ms increments) | 
| Arbitrary function |  | EasyArb | 
| Parameter |  | voltage, current, time | 
| Maximum number of points |  | 128 | 
| Dwell time |  | 10 ms to 600 s (10 ms increments) | 
| Repetition |  | continuous or burst mode with 1 to 255 repetitions | 
| Trigger | optional (R&S®NGE-K103) | manually, remote control or via optional trigger input | 
| Trigger and control interface | optional (R&S®NGE-K103) | digital I/O | 
| Minimum trigger interval |  | 10 ms | 
| Trigger response time |  | < 150 ms | 
| Trigger level |  | TTL, max. 5 V | 
| Edge direction |  | rising, falling | 
| Display and interfaces | 
| Display |  | 3.5"/QVGA | 
| Front panel connections channel outputs |  | 4 mm safety binding posts | 
| Remote control interfaces | Standard | USB-TMC, USB-CDC (virtual COM) | 
|  | optional (R&S®NGE-K101) | LAN (LXI) | 
|  | optional (R&S®NGE-K102) | wireless LAN 802.11 b/g/n, 2.4 GHz; operating mode: client | 
| Trigger and control interface | optional (R&S®NGE-K103) | digital I/O | 
| General data | 
| Environmental conditions |  |  | 
| Temperature | operating temperature range | 0°C to +40°C | 
|  | storage temperature range | –20°C to +70° | 
| Humidity | noncondensing | 5% to 80% | 
| Power rating |  |  | 
| Mains nominal voltage |  | 115 V/230 V (± 10%) | 
| Mains frequency |  | 50 Hz/60 Hz | 
| Maximum power consumption |  | 180 W | 
| Mains fuses | 115 V AC power source | IEC60127-2/5 T 5 A 250 V | 
|  | 230 V AC power source | IEC60127-2/5 T 2.5 A 250 V | 
| Product conformity |  |  | 
| Electromagnetic compatibility | EU, in line with EU RE Directive 2014/53/EU (notified body: CTC advanced GmbH) | applied standards: EN61326-1, EN61326-2-1,
 DRAFT EN301489-1 V 2.2.0,
 DRAFT EN301489-17 V 3.2.0,
 EN300328 V 2.1.1,
 47 CFR FCC part 15B (class A),
 and ICES-003 issue 6
 | 
|  | Korea | KC mark | 
| Electrical safety | EU, in line with Low Voltage Directive 2014/35/EU | EN61010-1 | 
|  | USA | UL 61010-1 | 
|  | Canada | CSA C22.2 No. 61010-1 | 
| Wireless LAN certifications | Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Kingdom | CE | 
|  | Singapore | iMDA standards DB102020 | 
|  | United States, Canada | FCC, IC | 
| RoHS |  | EN50581, in line with EU directive 2011/65/EU | 
| International safety approvals | cTUVus mark | certificate no. U8 18 04 87787 008 | 
| Mechanical resistance |  |  | 
| Vibration | Sinusoidal | 5 Hz to 55 Hz, 0.30 mm (peak-to-peak) amplitude const., 55 Hz to 155 Hz, 0.5 g const., in line with EN60068-2-6 | 
|  | Random | 8 Hz to 500 Hz, 1.2 g (RMS), in all three axes, in line with EN60068-2-64 | 
| Shock |  | 10 Hz to 45 Hz, ramp 6 dB/octave, 45 Hz to 2000 Hz: max. 40 g in line with MIL-STD-810E | 
| Mechanical data |  |  | 
| Dimensions | W × H × D | 222 mm × 97 mm × 310 mm (8.74 in × 3.82 in × 12.21 in)
 | 
| Weight | R&S®NGE102B | 4.9 kg (10.80 lb) | 
|  | R&S®NGE103B | 5.0 kg (11.00 lb) | 
| Rack installation | optional (R&S®HZC95) | ½ 19" | 
| Recommended calibration interval | operation 40 h/week in the full range of the specified environmental conditions | 1 year |