| Outputs | The channel outputs are galvanically isolated and not connected to ground. | 
| Number of output channels | R&S®NGP802, R&S®NGP822 | 2 | 
|  | R&S®NGP804, R&S®NGP824, R&S®NGP814 | 4 | 
| Total output power | R&S®NGP802, R&S®NGP822 | max. 400 W | 
|  | R&S®NGP804, R&S®NGP824, R&S®NGP814 | max. 800 W | 
| Maximum output power per channel | 200 W | 
| Output voltage per channel | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | 0 V to 32 V | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | 0 V to 64 V | 
| Maximum output current per channel | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | 20 A | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | 10 A | 
| Maximum voltage in serial operation | R&S®NGP802 | 64 V | 
|  | R&S®NGP822, R&S®NGP804, R&S®NGP814 | 128 V | 
|  | R&S®NGP824 | 250 V | 
| Maximum current in parallel operation | R&S®NGP822 | 20 A | 
|  | R&S®NGP802, R&S®NGP824, R&S®NGP814 | 40 A | 
|  | R&S®NGP804 | 80 A | 
| Voltage ripple and noise | 20 Hz to 20 MHz | < 3 mV (RMS), < 30 mV (Vpp) (meas.)
 | 
| Current ripple and noise | 20 Hz to 20 MHz | < 3.5 mA (RMS) (meas.) | 
| Load regulation | load change: 10% to 90% |  | 
| Voltage | ±(% of output + offset) |  | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | < 0.01% + 5 Mv | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | < 0.01% + 10 mV | 
| Current | ±(% of output + offset) | < 0.01% + 5 mA | 
| Load recovery time | 50% to 100% load change to within 0.2% of rated voltage | < 400 µs (meas.) | 
| Rise time | 10% to 90% of rated output voltage, resistive load | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | < 10 ms | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | < 12 ms | 
| Fall time | 90% to 10% of rated output voltage, resistive load | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | full load: < 10 ms, no load: < 50 ms | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | full load: < 25 ms, no load: < 50 ms | 
| Programming resolution | 
| Voltage |  | 1 mV | 
| Current |  | 0.5 mA | 
| Programming accuracy | 
| Voltage | ±(% of setting + offset) |  | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | < 0.05% + 5 mV | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | < 0.05% + 10 mV | 
| Current | ±(% of setting + offset) |  | 
|  |  | < 0.1% + 5 mA | 
| Output measurements | 
| Measurement functions | voltage, current, power, energy | 
| Readback resolution | 
| Voltage |  | 1 mV | 
| Current |  | 0.5 mA | 
| Readback accuracy | 
| Voltage | ±(% of output + offset) |  | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | < 0.05% + 5 mV | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | < 0.05% + 10 mV | 
| Current | ±(% of output + offset) |  | 
|  |  | < 0.1% + 5 mA | 
| Temperature coefficient (per °C) | ±(% of output + offset), +5°C to +20°C and +30°C to +40°C | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | voltage: < 0.0075% + 0.75 mV, current: < 0.015% + 0.75 mA | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | voltage: < 0.0075% + 1.5 mV, current: < 0.015% + 0.75 mA | 
| Remote sensing | 
| Maximum sense compensation | 1 V (meas.) | 
| Ratings | 
| Maximum voltage to ground | 250 V DC | 
| Maximum counter | voltage voltage with the same polarity connected to the outputs | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | 35 V | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | 70 V | 
| Maximum reverse | voltage voltage with opposite polarity connected to the outputs | 0.4 V | 
| Maximum reverse current | for 5 min max. | 20 A | 
| Remote control | 
| Command processing time | < 6 ms (typ.) | 
| Protection functions | 
| Overvoltage protection | adjustable for each channel | 
| Programming resolution | 1 mV | 
| Overpower protection | adjustable for each channel | 
| Overcurrent protection (electronic fuse) | adjustable for each channel | 
| Programming resolution | 0.5 mA | 
| Response time | (Iload > Iresp × 2) at Iload ≥ 2 A | < 1 ms | 
| Fuse linking (FuseLink function) | Yes | 
| Fuse delay at output-on | adjustable for each channel | 10 ms to 10 s (1 ms increments) | 
| Fuse delay time | adjustable for each channel | 10 ms to 10 s (1 ms increments) | 
| Response time for linked channels | < 5 ms | 
| Overtemperature protection | independent for each channel | 
| Special functions | 
| Output ramp function | EasyRamp | 
| EasyRamp time | 10 ms to 60 s (1 ms increments) | 
| Output delay |  | 
| Synchronicity | < 1 ms (typ.) | 
| Delay per channel | 10 ms to 10 s (1 ms increments) | 
| Arbitrary function | QuickArb | 
| Parameters | voltage, current, time | 
| Maximum number of points | 1024 | 
| Maximum number of subgroups | 8 | 
| Dwell time | 1 ms to 60 s (1 ms increments) | 
| Repetition | continuous or burst mode with 1 to 65535 repetitions | 
| Trigger | manually, by remote control or via optional trigger input | 
| Trigger and control interfaces | R&S®NGP-K103 | digital I/O, 16-pin connector block | 
| Trigger response time | < 3 ms (typ.) | 
| Maximum voltage (IN/OUT) | 5.5 V | 
| Input trigger level | TTL | 
| Maximum drain current (OUT) | 5 mA | 
| Analog control interface | R&S®NGP-K107 | analog input, 16-pin connector block | 
| Input voltage | 0% to 100% control of voltage or current | 0 V to 5 V | 
| Output accuracy | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | voltage: < 0.1% + 16 mV, current: < 0.1% + 30 mA
 | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | voltage: < 0.1% + 32 mV, current: < 0.1% + 15 mA
 | 
| Temperature coefficient | (per °C) ±(% of output + offset), +5°C to +20°C and +30°C to +40°C |  | 
|  | R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) | voltage: < 0.015% + 2.4 mV, current: < 0.015% + 4.5 mA
 | 
|  | R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) | voltage: < 0.015% + 4.8 mV, current: < 0.015% + 2.25 mA
 | 
| Update rate | 1 ms | 
| Data logging | 
| Maximum acquisition rate | 125 sample/s | 
| Memory depth | 800 Mbyte internal or external memory | 
| Voltage resolution | see readback resolution | 
| Voltage accuracy | see readback accuracy | 
| Current resolution | see readback resolution | 
| Current accuracy | see readback accuracy | 
| Display and interfaces | 
| Display | TFT 5" 800 × 480 pixel WVGA touch | 
| Front panel connections | 4 mm safety sockets (channel outputs, remote sensing)
 | 
| Rear panel connections | R&S®NGP802, R&S®NGP822 | 8-pin connector block (channel outputs and remote sensing)
 | 
|  | R&S®NGP804, R&S®NGP824, R&S®NGP814 | 2 × 8-pin connector block (channel outputs and remote sensing)
 | 
| Remote control interfaces | Standard | USB-TMC, USB-CDC (Virtual COM), LAN | 
|  | R&S®NGP-K102 | WLAN | 
|  | R&S®NG-B105 | IEEE-488 (GPIB) | 
| General data | 
| Environmental conditions | 
| Temperature | operating temperature range | +5°C to +40°C | 
|  | storage temperature range | –20°C to +70°C | 
| Humidity | Noncondensing | 5% to 95% | 
| Power rating | 
| Mains nominal voltage | 100 V to 250 V | 
| Mains frequency | 50 Hz to 60 Hz | 
| Maximum power consumption | R&S®NGP802, R&S®NGP822 | 650 W | 
|  | R&S®NGP804, R&S®NGP824, R&S®NGP814 | 1125 W | 
| Mains fuses | internal (not user accessible) | 16 A 250 V IEC60127-2/7 fast acting | 
| Product conformity | 
| Electromagnetic compatibility | EU: in line with Radio Equipment Directive 2014/53/EU | applied standards: ► ETSI EN300328 V2.1.1
 ► EN61326-1
 ► EN61326-2-1
 ► EN55011 (Class A)
 ► EN55032 (Class A)
 ► ETSI EN301489-1 V2.2.0
 ► ETSI EN301489-17 V3.2.0
 | 
|  | Korea | KC mark | 
|  | USA, Canada | FCC47 CFR Part 15B, ICES-003 Issue 6 | 
| Electrical safety | EU: in line with Low Voltage Directive 2014/35/EU | applied harmonized standards: EN61010-1 | 
|  | USA, Canada | UL61010-1, CSA C22.2 No. 61010-1 | 
| WLAN approvals | Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Kingdom | CE | 
|  | Singapore | iMDA standards DB102020 | 
|  | USA, Canada | FCC, IC | 
| RoHS | in line with EU Directive 2011/65/EU | EN50581 | 
| Mechanical resistance | 
| Vibration | Sinusoidal | 5 Hz to 55 Hz, 0.3 mm (peak-to-peak), 55 Hz to 150 Hz, 0.5 g const., in line with EN60068-2-6 | 
|  | Random | 8 Hz to 500 Hz, acceleration: 1.2 g (RMS), in line with EN60068-2-64 | 
| Shock |  | 40 g shock spectrum, in line with MIL-STD-810E, method 516.4, procedure I | 
| Mechanical data | 
| Dimensions | W × H × D | 362 mm × 100 mm × 451 mm (14.25 in × 3.94 in × 17.76 in)
 | 
| Weight | R&S®NGP802, R&S®NGP822 | 7.5 kg (16.5 lb) | 
|  | R&S®NGP804, R&S®NGP824, R&S®NGP814 | 8.0 kg (17.6 lb) | 
| Rack installation | R&S®ZZA-GE23 | 19 in, 2 HU | 
| Recommended calibration interval | operation 40 h/week over entire range of specified environmental conditions | 1 year |