    產品簡述
    ♦ 頻率範圍: 10 Hz ~ 44 GHz
    ♦ 解析頻寬: 200 MHz
    ♦ 相位雜訊: –114 dBc/Hz
    ♦ 支援多點觸控的圖形界面、SCPI 記錄器以及基於事件的操作
    ♦ 多種類比和數位信號分析選配,包括5G NR 信號分析
    ♦ 10 Gbit/s LAN 串口加快數據傳輸(選配)

      產品詳細介紹Product Introduction

      Measurements at the push of a button, capturing rare events with event based actions, and simple script programming with the SCPI recorder – setting up and performing complex measurements in no time is the strength of the R&S®FSV3000 signal and spectrum analyzer. Its fast measurement speed is a major asset especially in automated production applications as it yields high throughput.

      The R&S®FSV3000 offers functions that make the configu‑ ration of complex measurements fast and easy. Setting up RF parameters with touchscreen gestures is as easy as using your smartphone. The autoset feature automatically sets key parameters such as frequency, level and gating. An SCPI recorder, which translates manual operation into remote control command scripts, considerably speeds up script programming. And event based actions support you when debugging your DUT by capturing and documenting rare events.

      The R&S®FSV3000 has been designed for high measurement speed. This is a major advantage especially in pro‑ duction environments, which benefit from fast measurement times for spectral measurements, high-speed signal demodulation and rapid switching between different mea‑ surement modes. The R&S®FSV3000 provides digital modulation analysis up to 200 MHz analysis bandwidth for cellular and wire‑ less standards, including 5G NR.


      Available modelsR&S®FSV3004R&S®FSV3007R&S®FSV3013R&S®FSV3030R&S®FSV3044
      Frequency range10 Hz - 4 GHz10 Hz - 7.5 GHz10 Hz - 13.6 GHz10 Hz - 30 GHz10 Hz - 44 GHz
      Phase noise< –114 dBc (1 Hz)

      (f = 1 GHz, 10 kHz offset, option B710)
      < –114 dBc (1 Hz)

      (f = 1 GHz, 10 kHz offset, option B710)
      < –114 dBc (1 Hz)

      (f = 1 GHz, 10 kHz offset, option B710)
      < –114 dBc (1 Hz)

      (f = 1 GHz, 10 kHz offset, option B710)
      < –114 dBc (1 Hz)

      (f = 1 GHz, 10 kHz offset , option B710)
      DANL @ 1 GHz< –162 dBm/Hz< –162 dBm/Hz< –162 dBm/Hz< –162 dBm/Hz< –162 dBm/Hz
      Max. analysis bandwidth200 MHz200 MHz200 MHz200 MHz200 MHz

      Features
      • Frequency range from 10 Hz to 4 GHz/7.5 GHz/13.6 GHz/ 30 GHz/44 GHz
      • Frequency range extension up to 500 GHz with external harmonic mixer from Rohde&Schwarz
      • Frequency range extension up to 50 GHz for best performance signal analysis with external frontends from Rohde&Schwar
      • Analysis bandwidth up to 200 MHz
      • SSB phase noise at 10 kHz offset (1 GHz): < –114 dBc (1 Hz)
      • Third-order intercept (TOI) at 1 GHz: +18 dBm (typ.)
      • DANL at 1 GHz: –151 dBm
      • DANL at 1 GHz with optional preamplifier: –165 dBm
      • Ready for cloud based testing
      • 10 Gbit/s LAN interface (option)
      • User interface with multitouch display, SCPI recorder and event based actions
      • Measurement applications for analog and digital signal analysis, including 5G NR

      Specification
      Specifications in brief
      Frequency
      Frequency range R&S®FSV3004 10 Hz to 4 GHz
      R&S®FSV300710 Hz to 7.5 GHz
      R&S®FSV3013 10 Hz to 13.6 GHz
      R&S®FSV3030 10 Hz to 30 GHz
      R&S®FSV3044 10 Hz to 44 GHz
      Aging of frequency reference 1 · 10–6 per year
      with R&S®FSV3-B4 option 1 · 10–7 per year
      Bandwidthstandard filter 1 Hz to 10 MHz
      Resolution bandwidth RRC filter 18 kHz (NADC),
      24.3 kHz (TETRA),
      3.84 MHz (3GPP),
      4.096 MHz
      channel filter 100 Hz to 5 MHz
      video filter 1 Hz to 10 MHz
      I/Q demodulation bandwidth standard 28 MHz
      with R&S®FSV3-B40 option 40 MHz
      with R&S®FSV3-B200 option 200 MHz
      Phase noise (with R&S®FSV3-B710 option)1 GHz carrier
         1 kHz offset < –109 dBc (1 Hz)
         10 kHz offset < –114 dBc (1 Hz)
         100 kHz offset < –119 dBc (1 Hz)
         1 MHz offset < –135 dBc (1 Hz)
      Displayed average noise level (DANL) 1 GHz –151 dBm (typ.)
      DANL with preamplifier (R&S®FSV3-B24 option) 50 MHz ≤ f < 3 GHz –165 dBm (typ.)
      Intermodulation
      Third order intercept (TOI) 1 GHz > 15 dBm, 18 dBm (typ.)
      Total measurement uncertainty 2 GHz 0.29 dB